J.ophthalmol.(Ukraine).2015;5:18-24.

https://doi.org/10.31288/oftalmolzh201551824

Two-year results of treatment of open-angle glaucoma and concomitant cataract with combined phacoemulsification and Alcon EX-PRESS glaucoma filtration implant surgery

S. P. Medvedchuk,1 PhD candidate, P.A. Bezdetko,2 Dr. Sc. (Med), Prof., G.Ia. Parkhomenko,2 Cand. Sc. (Med)

1 "New Vision" Clinic Group, Cherkasy, Kyiv, Khmelnytsky, Dnipropetrovsk, Sumy, Ukraine,

2 Department of Ophthalmology, Kharkiv National Medical University

E-mail: ck-maindoctor@zir.org.ua

Background: A combined approach for glaucoma coexistent with cataract has been always interesting for ophthalmologists since these diseases are often found together in patients.

Purpose: To investigate the clinical efficacy and safety of combined phacoemulsification and Alcon EX-PRESS glaucoma filtration device (GFD) implant procedure and to develop indications for implantation of 50-?m and 200-?m devices.

Materials and Methods: The study involved 115 primary open-angle glaucoma patients (115 eyes) with cataract divided into 3 groups depending on the type of glaucoma surgery. Group 1 involved 30 patients (30 eyes) who underwent phacoemulsification combined with the 50-?m GFD implant procedure. Group 2 involved 30 patients (30 eyes) who underwent phacoemulsification combined with the 200-?m GFD implant procedure. Group 3 included 55 patients (55 eyes) who underwent phacoemulsification combined with nonperforating deep sclerectomy. The mean preoperative intraocular pressure (IOP) in Group 1, Group 2 and Group 3 was 22.9±2.9 mmHg, 22.8 ± 3.5 mmHg (р1/2 > 0.05), and 22.6 ± 3.4 mmHg (р1/3 > 0.05; р2/3 > 0.05), respectively.

Results: At 24 months postoperative, the mean IOP in Group 1, Group 2 and Group 3 was 15.4 ± 1.7 mmHg, 16.1 ± 2.1 mmHg (р1/2 > 0.,05), and 18.8 ± 1.9 mmHg (р1/3 < 0.05; р2/3 < 0.05), respectively. The number of postoperative complications in Group 1 and Group 2 was statistically significantly lower than that in Group 3.

Conclusion: Phacoemulsification combined with the EX-PRESS glaucoma filtration device implant procedure is an efficacious and safe surgical technique for stabilization of glaucoma process. The 50-?m Alcon EX-PRES GFD implantation is indicated in stage I or stage II of unstabilized glaucoma with subcompensated IOP. The 200-?m Alcon EX-PRESS GFD implantation is indicated in stage III or stage IV of unstabilized glaucoma with decompensated IOP.

Keywords: open-angle glaucoma, EX-PRESS glaucoma filtration device, intraocular pressure, cataract phacoemulsification.

References
1. Zbitneva SV. [Optimization of phacoemulsification treatment combined with non-perforating deep sclerectomy for cataract and glaucoma], [Abstract of Cand. Sc. (Med) thesis]. Kyiv: Shupik National Medical Academy of Continuing Education; 2008. Russian
2. Mogilevskii SIu. [Cataract with concomitant primary glaucoma (pathogenetically oriented surgery and laser treatment)]. [Abstract of Dr. Sc. (Med) thesis]. Donetsk: Gor’kii Donetsk Medical University; 2007. Russian 
3. Pavliuchenko KP, Mogilevskii SIu, Oleinik TV. [Efficacy of phacoemulsification, intraocular lens implantation and trabeculectomy in cataract patients with primary glaucoma]. Oftalmol Zh. 2005;2:13-7. Russian
4. Borras O, Echaque J. Study: non-perforating “Shlemmectomy” an effective glaucoma treatment option. Ocular Surg News. 2000;18(18): 24-5
5. Ceki? O, Batman C. Hyposecretion of aqueous: another mechanism for reduced intraocular pressure after phacoemulsification [letter; comment]. J Cataract Refract Surg. 1998 May;24(5):574
Crossref
6. Lachkar Y, Neverauskiene J, Jeanteur-Lunel MN et al. Nonpenetrating deep sclerectomy: a 6-year retrospective study. Eur J Ophthalmol. 2004 Jan-Feb;14(1):26-36
Crossref
7. Meyer MA, Savitt ML, Kopitas E. The effect of phacoemulsification on aqueous outflow facility. Ophthalmology. 1997 Aug;104(8):1221-7
Crossref
8. Neroev VV, Bykov VP, Kvasha OI, Belyovtseva TA. [Glaucoma micro drainage surgery: literature review]. Klin. oftalmologiia 2009;3:113-6. Russian
9. Nesterov AP. Glaucoma. Moscow: Meditsina; 1995.  
10. Kuroedov AV, Ogorodnikova VIu. [Late results of use of EX-PRESS glaucoma drainage device in patients with advanced stages of primary open-angle glaucoma]. Oftalmologiia 2012;1:38-42. Russian
11. de Jong L, Lafuma A, Aguad? AS, et al. Five-year extension of a clinical trial comparing the EX-PRESS glaucoma filtration device and trabeculectomy in primary open-angle glaucoma. Clin Ophthalmol. 2011;5:527-33
Crossref
12. Alekseev BN, Pisetskaia SF. [Ocular hydro- and hemodynamics in detachment of the ciliary body or choroid following glaucoma surgery]. Oftalmol Zh. 1983;6:347-49. Russian
13. Chung AN, Aung T, Wang JC, Chew PT. Surgical outcomes of combined phacoemulsification and glaucoma drainage implant surgery for Asian patients with refractory glaucoma with cataract. Am J Ophthalmol. 2004 Feb;137(2):294-300
Crossref
14. Jurowski P, Go? R. Effectiveness of the combined surgical treatment for glaucoma and cataract. Klin Oczna. 2005;107(4-6):212-6
15. Ku WC, Lin YH, Chuang LH, Yang KJ. Choroidal detachment after filtering surgery. Chang Gung Med J. 2005 Mar;28(3):151-8
16. Blumenthal M. Manual ECCE, the present state of the art. Klin Monbl Augenheilkd. 1994 Nov;205(5):266-70
Crossref
17. Bloomberg L.B. Modified trabeculectomy / trabeculotomy with no-stitch cataract surgery. J Cataract Refract Surg. 1996 Jan-Feb;22(1):14-22
Crossref
18. Munden PM, Alward WL. Combined phacoemulsification, posterior chamber intraocular lens implantation, and trabeculectomy with mitomycin C. Am J Ophthalmol. 1995 Jan;119(1):20-9
Crossref
19. Zueva MV, Sheludchenko VM, Shpak AA. [Up-to-date ocular diagnostic techniques: summary and prospects]. In: [Proceedings of the 9th Russian Ophthalmological Congress]; Moscow (Russia): 2010. p. 498-500. Russian
20. Egorov EA, Kurmangalieva MM, Fefotovskikh GV. [Morphological changes in the retina of glaucoma patients]. Klin. oftalmologiia 2004;5(2):54-6. Russian
21. Mosin IM. [Optical coherent tomography]. In: Shamshinova AM, ed. [Essays on clinical ocular physiology]. 3rd ed. Moscow: Nauchno-meditsinskaia firma MBN; 2006. p.785-658